Many excellent reference texts and booklets are currently available. The
following is a brief listing:

Laboratory Ventilation Standards

Federal Register 29 CFR Part 1910
Non-mandatory recommendations from “Prudent Practices.”
• Fume hoods should have a continuous monitoring device

• Face velocities should be between 60-100 linear feet per minute (lfpm)

• Average 2.5 linear feet of hood space per person

Occupational Health and Safety
U.S. Department of Labor
200 Constitution Avenue N.W.
Washington, DC 20210
(202) 523-1452


Industrial Ventilation-ACGIH
• Fume hood face velocities between 60-100 lfpm

• Maximum of 125 lfpm for radioisotope hoods

• Duct velocities of 1000-2000 fpm for vapors, gasses and smoke

• Stack discharge height 1.3-2.0 x building height

• Well designed fume hood containment loss, <0.10 ppm