Chapter 6: Maintaining Your Protector XLE

Product Service 1-800-522-7658

24

• The hood baffles should be checked for blockages behind them to ensure

that the hood is maintaining proper airflow.

• All weekly activities.


Annually

• Replace the fluorescent lamps.

• Have the fume hood recertified by a qualified certification technician. See

Certifying the Protector XLE Fume Hood in Chapter 3.

• All monthly activities.


Biannually

• The sash assembly should be checked to ensure proper operation and to

make sure there are no signs of abnormal wear on the sash pulleys, cables
and clamps.

Routine Service Operations


Front Panel Removal:

1. Simply lift the front panel up and then away from the hood to provide access

to the top.


Changing the Fluorescent Lamp:

1. Turn light switch to “OFF.”
2. Remove the front panel as noted earlier.
3. Reach over the front header of the hood and remove the knock out plugs at

both ends of fixture. Lift fixture up.

4. Remove the fluorescent lamp by pushing it out of the spring-loaded lamp

socket and swinging it out of the other lamp socket.

5. Install the new lamp by reversing the removal procedure.

Your Next Step


After you understand the maintenance procedures, you are ready to proceed to
Chapter 7: Modifying Your Protector XLE.

S

A