MAINTENANCE
Attaching the Right Arm
Open the torso as described in steps 1 - 4, Opening the Torso.
Follow the procedures listed below in reverse to detach the arms.
1 Align the arm axle with the shoulder socket.
2 Feed the cables from the arm axle through the shoulder socket
3 Push the axle into the bracket until it is flush with the inside of
the bracket.
4 Tighten the shoulder screw with the Allen wrench.
JAW
ANT RIGH
T
ARM
LA CONV
RIGH
ARMRM
LA CONV
ANT
RIG
A
AWAN
L
JA
5 Connect the corresponding cables as shown below.
Right Arm to Torso – Cable and Tube Descriptions
Name/Label
Tube/Cable Color
Connector Description
Right Arm
Black harness cable
Black rectangular connector,
4 lead
Air
Transparent Tube
White tube twist connector
Right Radial
Grey harness cable
Black with silver colored
connector
Replacing Crico Tape/Neck Skin
After creating an emergency airway through the cricothyroid
membrane, replace the perforated membrane before starting a new
simulation session.
1 Remove the Neck Skin (velcro fasteners behind the neck).
2 Remove the old strip of Crico Tape.
3 Replace with a new Crico Tape.
4 Ensure Crico Tape completely covers and seals the opening to
prevent leakage while ventilating the Patient Simulator.