35

MAINTENANCE

Attaching the Right Arm

Open the torso as described in steps 1 - 4, Opening the Torso.
Follow the procedures listed below in reverse to detach the arms.

1 Align the arm axle with the shoulder socket.

2 Feed the cables from the arm axle through the shoulder socket

3 Push the axle into the bracket until it is flush with the inside of

the bracket.

4 Tighten the shoulder screw with the Allen wrench.

JAW

ANT RIGH

T

ARM

LA CONV

RIGH

ARMRM

LA CONV

ANT

RIG

A

AWAN

L

JA

5 Connect the corresponding cables as shown below.

Right Arm to Torso – Cable and Tube Descriptions

Name/Label

Tube/Cable Color

Connector Description

Right Arm

Black harness cable

Black rectangular connector,
4 lead

Air

Transparent Tube

White tube twist connector

Right Radial

Grey harness cable

Black with silver colored
connector

Replacing Crico Tape/Neck Skin

After creating an emergency airway through the cricothyroid
membrane, replace the perforated membrane before starting a new
simulation session.

1 Remove the Neck Skin (velcro fasteners behind the neck).

2 Remove the old strip of Crico Tape.

3 Replace with a new Crico Tape.

4 Ensure Crico Tape completely covers and seals the opening to

prevent leakage while ventilating the Patient Simulator.