Troubleshooting, Airway contamination, Chest movement – Laerdal SimMan Essential Bleeding User Manual

Page 40: Blood system, Troubleshooting the iv arm

Advertising
background image

40

Airway Contamination

Problem

– Simulator airways have become contaminated from mouth-to-

mouth rescue breathing

Solution

– Clean the outside of the Patient Simulator with Manikin Wipes.

Clean the inside of the oral cavity with Manikin Wipes. Change
the Lung Bladders; see: Maintenance - Replacing Lung Bladders.

Note: The Patient Simulator airways are not designed for

mouth to mouth rescue breathing or to be disinfected.

Chest Movement

Problem

– No chest rise on Patient Simulator.

Solution

– Check that Patient Simulator is switched ON.

– Check that Patient Simulator is not in sleep mode due to

inactivity. Reactivate the Patient Simulator.

– Check that awRR is not set to zero in LLEAP (Patient Case or

scenario)

– Check that airway complications like maximum air resistance or

laryngospasm are not set.

– Check if the internal compressor is switched off. See: Turning the

Internal Compressor Off/On.

– Check that any external compressed air source is switched off

and that the air tube is disconnected from the Patient Simulator.

– The internal compressor may have overheated. Wait

approximately 20 minutes for it to cool down. Remove blankets
or covers from the Patient Simulator, open the Torso Skin to
facilitate cooling.

– Chest Rise is set to bilateral (for example if ET-tube is inserted

too far into the bronchia).

– Chest Rise Bladder is leaking or tubing to Chest Rise Bladder is

twisted, kinked or disconnected. Replace Chest Rise Bladder if it
is leaking, see Maintenance: Replacing Chest Rise Bladders.

– Check air tubing for leakage; check that all connections are intact.

Replacing tubing if necessary.

– Shallow chest movements and the internal compressor runs

continuously. The internal compressor may be worn - contact
Laerdal Technical Service.

Blood System

Problem

– No blood flow.

Solution

– Make sure that the internal reservoir is filled with fluid.

Problem

– Too low flow.

Solution

– Clean the blood system.

– Check the flow setting in LLEAP.

– The blood system may need to be re-calibrated. Contact your

local service representative.

– Replace filter in pelvis.

Problem

– Air when bleeding

Solution

– Bleed the blood reservoir empty, then refill.

Problem

– Simulated blood is leaking from the back of the right leg during

filling or when the Patient Simulator is turned off.

Solution

– The internal reservoir may need to be replaced. Contact your

local service representative.

Problem

– Can’t fill the blood reservoir.

Solution

– The filter inside the fill bottle may be clogged. If so, order a new

fill bottle

Troubleshooting the IV Arm

Problem

– Veins do not energize.

Solution:

– Ensure Patient Simulator is powered on and IV Arm cable and air

tube are connected to the Patient Simulator.

– Ensure the blood reservoir is filled, see Filling and priming the

system.

– Ensure the tourniquet area is activated (depressed).

– Look for a leak between the IV Arm pad and the IV Arm chassis.

– If the tourniquet area has been activated several times within a

short period of time, internal pressure in the IV Arm may drop.
Wait 30 seconds to let pressure rebuild and try again.

Problem

– When injecting more than 40ml fluid and prime switch is

depressed, no fluid is visible in the drain tube.

Solution

– Leak inside IV Arm. Contact Laerdal Technical Service.

TROUBLESHOOTING

Advertising