Troubleshooting

System Setup

Problem

– Lost data or total system failure (General System failure).

Solution

– Should system shutdown or all data is lost or corrupted; please

contact your local Laerdal Service Center.

Stability and Connection in Access Point Mode

(Ap) (Patient Simulator Created Network)

Problem

– Connection between LLEAP and/or Patient Monitor and Patient

Simulator is lost.

Causes

– Other Software on PCs may interfere with LLEAP or Patient

Monitor.

– Multiple network connections may interfere with transmissions

between our Software and Patient Simulator.

Possible Solutions

– Remove unnecessary Software on PCs.

– Disable other networks.

Stability and Connection In Client Mode

Problem

– Connection between LLEAP and Patient Monitor and Patient

Simulator is lost.

Causes

– Patient Simulator has been positioned where limited connectivity

to external network exists.

– Other Software on PCs may interfere with LLEAP and/or Patient

Monitor.

Possible solutions

– Move Patient Simulator to improve connectivity to external

network.

– Remove unnecessary Software on PCs.

LLEAP

Problem

– Unable to log on to the computer.

Causes

– If the user decides to change the log on behavior it will require

a password.

Solutions

– Default password is “SimUser”.


Changing Patient Simulator Wireless

Network Display name

If operating more than one Patient Simulator, ensure that each
simulator has its own unique SSID name. For more information see
Simulator Firmware & Network Wizard Help.

Voice Conference Application

Problem

– Sound from Patient Simulator to LLEAP is not working.

Possible Solutions

– Check that the microphone has not slipped out of place. Open

the head skin zipper, located at the back of the head, enough to
reveal the ears.

– Ensure that the microphone is positioned in the cup located

towards the top of the ear with the black surface facing
outwards.

Problem

– Instructor microphone not picking up sound.

Possible Solutions

– Unplug and re-plug the headset to your computer

– Ensure correct sound device is selected. In Voice Conference

Application main menu, select <Options>, <Select Device>.

– Check volume settings in Windows. Ensure microphone is not

muted.

Debriefing

See section Troubleshooting in SimView User Guide or
SessionViewer User Guide.

Patient Simulator

Problem

– Unpredictable behavior

Possible Solutions

– Patient Simulator malfunctions may be caused by loose cables,

tubes or connectors. Open the torso and check if any items
appear to have become disconnected or are leaking. See
Opening the Torso.

– In case of fluid leakage, power off the Patient Simulator and

contact Laerdal Technical Service.

Problem

– Identifying a single Patient Simulator when there are multiple

simulators on a network?

Solutions

– When checking the Patient Simulator’s pulse, the <Select

Simulator> dialog of LLEAP will indicate which Patient Simulator
is pulse palpated.

TROUBLESHOOTING