6 vacuum pump – Martin Christ Beta 2-8 LSCplus User Manual

Page 34

Advertising
background image

Beta 1-8 LSC&
Beta 2-8 LSC&

5 Set-up and connection

34

Version 04/2013, Rev. 1.1 of 16/12/2013 • sb

Translation of the original operating manual

NOTE

The vacuum sensor comes supplied in a calibrated state.


After the freeze-dryer has been switched on, the vacuum sensor needs
several minutes until it is ready for operation.

Pos: 104 /01 0 Univ ersal mod ule/ Sei tenwe chsel @ 0\ mod _12 021 1624 431 2_0 .docx @ 1 05 @ @ 1

Pos: 105 /20 0 Ch rist/3 61 G T-BA L abo r-Pilot (PROJEKTE)/Bet a 1 -8_ 2-8 LSCplus/ 050 Au fstellu ng u nd Ans chluss/ 050 -00 60 V akuu mpu mp e Beta @ 16 \mo d_1 359 106 6219 62_ 68. docx @ 80 814 @ 2 @ 1

5.4.6

Vakuum pum pe

5.6 Vacuum pump

NOTE

Please refer to the separate operating manual of the vacuum pump and
the exhaust filter!


The vacuum pump must be connected to the vacuum connector of the unit
and its plug must be plugged into the IEC C14 connector on the back of the
unit (see chapter 2.1.1 - "Functional and operating elements").
The oil mist that escapes when the pump is in operation must be retained
or carried off by way of an oil mist separator.
• We strongly recommend using an oil mist separator. This filter prevents

air pollution by oil mist.

• In order to carry off the oil mist, connect a suitable hose to the exhaust

port of the vacuum pump (RZ-2.5 and RC-6: ½", DUO 5 or DUO 10 ¾").

• The hose must be laid so that the condensate cannot flow back into the

pump. In the case of upward leading hoses, we recommend using a
separator (Woulfe's bottle or wash bottle).