Martin Christ Beta 2-8 LSCplus User Manual

Page 96

Advertising
background image

Beta 1-8 LSC&
Beta 2-8 LSC&

11 Appendix

96

Version 04/2013, Rev. 1.1 of 16/12/2013 • sb

Translation of the original operating manual

4 Ensure that all of the valves are closed.
5 Let the vacuum pump warm up 20 to 30 minutes before the freeze-

drying processes commences.

6 Connect a frozen sample to a valve.

CAUTION

After the pressure has fallen below 1.030 mbar, a frozen sample can be
connected to a valve. The next frozen sample cannot be connected to
another valve until the pressure is again lower than 1.030 bar.

NOTE

After the pressure has fallen below 1.030 mbar, a frozen sample can be
connected to a valve. The next frozen sample cannot be connected to
another valve until the pressure is again lower than 1.030 bar.


7 The operating panel displays the vacuum, the ice condenser

temperature, and the current operating mode.

8 The end of the process is reached when the ice condenser is no longer

loaded and when it again reaches a final temperature of approximately -
50°C to -54°C. The pressure decreases as a function of the ice
condenser temperature.

NOTE

The drying time depends on the layer thickness of the sample, the solids
content of the sample, and the amount of heat that is supplied during the
drying process. In the case of a layer thickness of 1 cm, the freeze-drying
process usually takes 24 hours.


9 Switch the vacuum pump off and aerate the drying chamber via the

condensate drain valve or via a rubber valve.

10 Switch the unit off by actuating the mains power switch and take the

product out of the freeze-dryer.

11 Switch the unit on again and start the defrosting process (button

"Operating mode: select/start" – "Defrosting").

CAUTION

Ensure that no water gets into the pipe connection of the vacuum pump or
vacuum sensor.


12 Drain the defrosting water via the defrosting water valve on the left-hand

side of the unit. To do so, connect a hose to the hose connector and
collect the defrosting water in a suitable vessel.

13 Keep the freeze-dryer open (i.e. without the lid or drying chamber) when

it is not in use so that moisture can evaporate. This increases the
service life of the vacuum sensor.