Version 04/2013, Rev. 1.1 of 16/12/2013 • sb

Translation of the original operating manual

• In order to assign an address, open the dialog box "Wireless Shelf

Technology".

• Ensure that only the WST module to which an address is to be assigned

is connected. If several modules are connected, they will all receive the
same address, resulting in a communication conflict.

• Press the button "Address assignment", process the prompt, and

confirm the address assignment.


Vacuum pump oil change (4)
The system monitors the oil change interval of the vacuum pump.
The interval can be adapted to the vacuum pump model and utilisation.
When the end of an oil change interval is reached, a corresponding
message will be displayed.
• Acknowledge the message.
• Change the oil of the vacuum pump.
• Reset the operating hour counter in the dialog box “Vacuum pump oil

change” by pressing the “reset” button.

Fig. 50: Dialog box "Vacuum pump oil change"


Maintenance (5)
The maintenance interval of the unit is fixed at 3,000 operating hours or at
least one maintenance per year.
When the end of a maintenance interval is reached, a corresponding
message will be displayed.
• Acknowledge the message.
• Make an appointment for the maintenance of your freeze-dryer (see

section 8 "Maintenance and service").

• After the maintenance, our service engineer will reset the operating

hour counter in the dialog box "Maintenance".