7 Malfunctions and error correction

Version 04/2013, Rev. 1.1 of 16/12/2013 • sb

75

Translation of the original operating manual

Type of error

Possible reason

Correction

The displayed vacuum value is not
correct

• Incorrect calibration


• The vacuum sensor is soiled

(e.g. due to water residues)

• The vacuum sensor is

defective.

• Calibrate the vacuum sensor

(see the separate operating
instructions of the vacuum

sensor).

• Clean the vacuum sensor.
• Check the vacuum display with

the aid of a reference device (if

available).

• see chapter 7.1.2.5 - "Vacuum

sensor"

The vacuum pump is not activated

• See the separate operating

instructions of the vacuum

pump.

• See the separate operating

instructions of the vacuum

pump.

Insufficient ice condenser or shelf

temperature

• The overpressure switch of the

refrigeration unit has tripped.

• The thermal circuit breaker has

tripped.

• Let the unit cool down.

• Ensure sufficient air circulation

(see chapter 7.1.3 - "Insufficient

ice condenser temperature")

NOTE

If it is impossible to eliminate the errors, contact the Christ service
department!

Pos: 162 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1

Pos: 163 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1

Pos: 164 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 70 Stö run gen und Fe hlers uche /07 0-0 010 -00 10 N etzau sfall @ 6\m od_ 129 913 751 4747 _68 .docx @ 3 5128 @ 3 @ 1

7.1.1 Power failure

The control system continues with the process after a power failure. The
preselected conditions remain saved even during a process run.
In the event of a power failure in the drying phase, the batch may become
unusable. Whether the batch can be saved or not depends on the drying
phase in which the product was when the power failure occurred.
• In the final drying phase, the product has reached a residual moisture

content of approx. 5%. Below this value, the product is generally not
damaged even if the power failure lasts for a longer period of time.

• If the product is in the main drying phase, we recommend aerating the

unit, removing the product, and storing it in a deep-freeze. The
defrosted condensate must be drained off prior to the next start.