Creating a blank program (1)
This button is used to call up a blank program template. Only section 1 is
pre-defined as the loading section. In this phase, the start conditions of the
programs are defined. The room temperature (20°C) is the standard
default.

NOTE

If the product is frozen outside the freeze-dryer (double-chamber method),
the shelf temperature must be adapted accordingly in section 1.


For all the other sections, the set values must be defined (see figure
below):
• Press the button “Insert section” (4) and select the position of the

section as well as the freeze-drying phase. The section will be inserted
at the defined position.

• Adapt the parameters of the inserted section. All of the values are not

available in all of the freeze-drying phases.

• Insert and edit the next section.

NOTE

During the creation of a program, the order of the individual freeze-drying
phases must be maintained. This means that it is not possible to insert a
freezing section after a main drying section.

“Program” buttons:
1 "Program"
2 "Program name"
3 "Show diagram "
4 "Insert section "
5 "Delete section "
6 Button for calling up

more functions and

set values

7 "Save"

Fig. 61: Editing a blank program template


• The button “Delete section” (5) can be used to delete sections.
• The button ">>more" (6) can be used to enter more functions and set

values in a program-related manner, depending on the type of freeze-
dryer (see the sections below).

• Proceed in this manner in order to create an entire program according

to your specific needs.