Display calibration – SpectraCal madVR User Manual

Page 2

Advertising
background image

17544 Midvale Ave. N. Shoreline, WA 98133 USA

Corporate: +1 206 420 7514 | Tech Support: +1 206 457 8949 |

www.spectracal.com

2.

Load the Color Cube workflow - in the CalMAN main menu, select “Open Workflow Template,”
then select “Color Cube (3D LUT).”

3.

If CalMAN is running on the display to be calibrated, make the Test Pattern Generator window
as small as possible, while still providing a full target for your color meter.

Display Calibration

Follow this procedure to:

a. Create a display calibration 3D LUT file for a video display that is driven from madVR.

b. Load the 3D LUT file(s) into madVR.

c. Validate the calibrated display performance.

Create Display Calibration 3D LUT

Proceed through each step of the Color Cube workflow in the normal fashion, using the following notes
for madVR specifics.

1. Initial Setup section
In the ‘Initial Setup’ section of the Color Cube workflow, step through the
following pages:

a. Session Setup - Meter Connect

1. Connect your color meter to the CalMAN computer.
2. On the Session Setup page, click the Find Meter button.
3. If your meter is listed on the Find Meters dialog, select it, then

click Search. If your meter is not listed on the dialog, just click
Search.

4. Position the meter on the target display.

b. Session Setup - Source Connect

1. On the CalMAN Session Setup page, click the Find Source button.
2. On the Source dialog (see below), under Manufacturer, select

“madVR.”

3. Under Model, select “madVR – madVR Test Pattern Generator.”
4. Enter the IP address of the target computer running madVR (if CalMAN is running on the

same computer as madVR, enter the default local IP of 127.0.0.1).

5. Click Connect.