Manual operation 3.1 run mode, 2 flow restriction – Casella CEL Apex personal sampling pumps User Manual

Page 16

Advertising
background image

16

ESP

Spanish,

DAN

Danish.

TEMP Sets the temperature display units asCentigrade or Fahrenheit.
VOL

Sets the volume display mode to Auto or m

3

Auto displays volumes below 1000 litres in litres and above this as m

3

.

m

3

always displays volumes in m

3

.

BUZZEr

Sets an audible alarm to:

OFF

No alarm,

Err

Any error sounds the alarm,

On

Any error or key press sounds

the alarm.

PROG Mode

Gain access to program features, including

programmable run times on the standard unit and all programmable features on the Apex Pro.

ON

Makes any advanced program features available,

OFF

Hides any advanced program features.

LIFE

Read-only message that shows the total run time for the pump. It can be be used to
determine maintenance schedules.

LCD

Sets the LCD contrast.

Use

to increase the value (make darker) and

to decrease (make lighter).

CAL SET

Activates Calibration Mode, which allows the entire calibration for the pump to be reset,
based on a two point calibration.

The instrument will have been accurately calibrated at the factory prior to delivery.
Therefore it is recommended that this operation be performed only as part of a routine
service or when specific fine tuning is required to ensure that any selected flow rate will
be as close to the true flow as possible.


NO

Do not recalibrate the pump.

YES

Recalibrate the pump.

For information about recalibrating the whole range of an Apex or Apex Pro unit, refer to Chapter 5.

3.

Use

or

to make a change, then press

to accept it and move to the next

option.

3.

MANUAL OPERATION


3.1

Run Mode

1.

To start sampling and enter Run Mode, press

while in Hold Mode.

In Run Mode the display will cycle between, current duration of the sample, volume sampled, and
internal air temperature.

2.

To stop sampling while in Run Mode, press and hold

until the

symbol stops blinking and

disappears.

The instrument will display and save the accumulated duration and sample values until manually
reset.
Clearing values by a manual reset to close the run sample upon completion of an event ensures
that all downloaded data will be conveniently arranged to correspond with the sample result.

3.2

Flow Restriction

If the input flow becomes restricted, both

and the Red LED will flash to indicate a problem and if

enabled, the buzzer will sound. When the problem lasts for eight or more seconds, the pump motor will
switch itself OFF. After one minute, the pump will switch ON again in an attempt to determine whether the
problem has gone. When the problem remains for a further eight seconds, the pump will switch OFF and

remain off with the instrument still ON and

displayed.

If the output flow is restricted, both

and the Red LED will flash to indicate a problem and if enabled,

Advertising
This manual is related to the following products: