10 minutes, During 8 hour shift – Casella CEL Apex personal sampling pumps User Manual

Page 19

Advertising
background image

19

For example, when 10 litres must be sampled over an 8-hour work shift, the pump flow can be set to run
at a flow rate of:

= 0.021 litres / min = 21 millilitres / min

10

8 x 60

For the whole shift, which is a very low flow.


On

1 min

On

1 min

On

1 min

On

1 min

On

1 min

On

1 min

On

1 min

On

1 min

On

1 min

On

1 min

02011

Run: 480 min

Exp: 10 min

TWA: 2%

Off

47 min

Off

47 min

Off

47 min

Off

47 min

Off

47 min

Off

47 min

Off

47 min

Off

47 min

Off

47 min

Off

47 min



Alternatively, the TWA mode can be used to run the pump for short periods at a higher rate, for example 1
l/min, with the pump actually running for only:

= 10 minutes

during 8 hour shift

10

1

Using TWA mode, the sampling (RUN) time should be set to 8 hours

and the pump-on exposure (EXP) time to 10 minutes, as follows.
1.

Enter Configuration Mode, select Program Mode / ON and accept it

as described in Section 2.4.

The instrument reverts to Hold Mode.

2.

Press

or

several times until TWA blinks on the display.

3.

Press to select it.
RUN is displayed and the run time blinks to indicate that it can be
changed.

4.

Use

or

to select a run time.

Sampling (RUN) durations between 3 minute and 1 hour can be set
in 1-minute steps and durations between 1 and 25 hours in 10-minute steps.

5.

Press to accept the selected run time.
EXP is displayed, and the exposure time blinks to show it can be
changed.

6.

Use

or

to select an exposure time.

The available range is from 2 minutes to “1 minute less than the run
time“, in 1-minute steps.

7.

Press

to accept an exposure time and start sampling.

In addition to the usual

and

symbols, the display shows PROG. to indicate that a program

is active.
The pump ON-times are fixed at 1 minute, and the instrument calculates the necessary OFF-times
to spread these 1 minute intervals evenly throughout the run time.
While the program is running, the display cycles through current accumulated run time, volume
sampled, air temperature and calculated TWA (as a percentage of run time).

During OFF-times (pump not running) the

and

symbols are displayed to show the

instrument is paused.

Once the run is completed, the pump will switch OFF and the display show the

symbol and a

PROGEnd message.

Figure 10: TWA ON/OFF cycle