Test sequence – AMETEK CTS 3.0 User Manual
Page 105
Advertising

User Manual
Compliance Test System 3.0
101
8.4
Test Sequence
The IEC 61000-4-13 standard specifies the following test flow diagram.
Start - Class 1/2
Determine appropriate test
configuration
Perform 8.2.1 Test
"Harmonic combination"
Flatcurve and Overswing
Any functional
anomalies?
No
No
Class 2
required?
'Meister
curve' required?
Any functional
anomalies?
Any functional
anomalies?
Any functional
anomalies?
Test Completed
FAIL
Record Results
Test Completed
PASS
Record Results
Yes
Yes
No
Perform 8.2.2 Test
"Frequency Sweep"
Perform 8.2.4 Test
"Meister Curve"
Perform 8.2.3 Test
"Individual Harmonics/
Interharmonics"
Yes
Yes
No
Yes
Yes
No
No
Figure 8-2: IEC 61000-4-13 Flow Chart Class 1 and 2.