139
Measurement
Specification
Unit
Power
Range PACS-1 / PACS-3
0.1 - 12,500
W/ph
Range PACS-3-75
0.1 - 22,500
W/ph
Accuracy
± 0.25 % ± 0.25 % FS ± 20mW
mW
Resolution
0.1
W
Apparent Power
Range PACS-1 / PACS-3
0.01 - 12,500
VA/ph
Range PACS-3-75
0.01 - 22,500
VA/ph
Accuracy
± 0.15% ± 0.15% FS ± 20
mVA
Resolution
0.01
VA
Power Factor
Range
0.00 - ± 1.00
Accuracy
± 0.05
Resolution
0.01
Crest Factor
Range
20:1
Accuracy
± 0.005
Resolution
0.001
Frequency
Range
45 - 65
Hz
Accuracy
0.1 % of reading
Hz
Resolution
0.1
Hz
Harmonic Analysis
Range
Fundamental to 40
th
Accuracy Fundamental
±0.05% FS±0.05%/kHz
Accuracy Harmonics
±0.1 %±0.1%/kHz
Measurement window
10, 12 or 16 periods
Smoothing filter
1.5
sec
Flicker
Pst
Range
0.1 - 10
Pst
Accuracy
3
%
Resolution
0.01
Integration time
10
min
Plt
Range
0.1 - 10
d
max
Range
0 - 100
%
d
c
Range
0.1 - 100
%
d
t
Range
0.1 - 100
%
d
t
over3%(4%)
Range
0 - 1000
ms