Environmental, Regulatory – AMETEK CTS 3.0 User Manual
Page 145

User Manual
Compliance Test System 3.0
141
Frequency may not be 50Hz depending on power source used and
application.
15.4
Environmental
Parameter
Specification
Operating Temp:
0 degrees to +40 degrees Celsius.
Storage Temp:
0 degrees to +70 degrees Celsius.
Humidity:
Operating:.
90% RH up to 40 C
Storage:
90% RH up to 40 C, 75% RH up to 70 C.
Creepage and
Clearance:
Rated for Pollution Degree 2.
Insulation:
Rated to Installation Category(Overvoltage Category) II
Vibration:
Designed to meet NSTA 1A transportation levels.
Shock:
Designed to meet NSTA 1A transportation levels.
15.5
Regulatory
Parameter
Specification
Electromagnetic Emissions and
Immunity:
Designed to meet EN50081-1 and EN50082-1
European Emissions and Immunity standards as
required for the “CE” mark.
Acoustic Noise:
60 dBA maximum. Measured at one meter. (PACS-X
or 1251RP)
75 dBA maximum. Measured at one meter. (5001iX)
Safety:
Designed to meet EN61010-1 European safety
standards as required for the “CE” mark.