10.5 Output Subsystem
This subsystem controls the main output function of the RP Series source.
Subsystem Syntax
OUTPut
[:RELay]
Opens or closes the output relay.
DROP
Drops the output voltage for a specified amount of time or
indefinitely.
STARt:
[STATe]
Enables or disables the start phase angle mode
PHASe
Sets the start phase angle value in degrees.
OUTPut[:RELay]
The output of the AC source can be isolated from the unit under test by opening the output
relay. The programmed output voltage is set to zero before opening the relay and is returned
to its programmed value when closing the relay. This avoids hot switching the output relay
contacts.
Command Syntax
OUTPut[:RELay] <bool>
Parameters
<Bool> 0 | 1
*RST Value
defined by the PONS
Examples
OUTP 1
Query Syntax
OUTPut?
Returned Parameters
0 | 1
Related Commands
OUTP:DROP
OUTP:STAR