10.2.3 Output Calibration
The AC voltage output calibration is performed by adjusting the output up or down using a
calibration coefficient. The coefficient ranges from -127 (adjust upward) to + 127 (adjust
downward). An external traceable reference AC voltmeter should be used for this purpose.
CALibrate[:SOURce]:VOLTage <NRf>
This command will set the calibration coefficient for the AC full scale output voltage.
Command Syntax
CALibrate[:SOURce]:VOLTage <NRf>
Parameters
<NRf> (a value between -127 and +127)
Examples
CAL:VOLT -2
Query Syntax
CALibrate:VOLTage?
Returned Parameters
<NR1> (value range -127 to +127)
Related Commands
CAL:STAT