2 measurements [option – AMETEK 2001RP User Manual
Page 15

User and Programming Manual - Rev R
2001RP
5
2.2 Measurements [Option]
Parameter
Specification
Unit
True RMS Current
Available on standard unit.
Ranges
0.000 - 4.000
0.00 – 18.00
A
RMS
A
RMS
Accuracy
0.2 % FS + 0.3 % reading
Resolution
0.001
0.01
A
RMS
A
RMS
Temp
Coefficients
0.016
% FS /
°C
Peak Current
Ranges
(Remote only) 0.00 – 12.00
0.0 – 70.0
A
PEAK
A
PEAK
Accuracy
0.5 % FS + 0.5 % reading
Resolution
0.01
0.1
A
PEAK
A
PEAK
Crest Factor
Available over bus only. Requires interface option
Range
0.00 - 4.00
Accuracy
0.05
Resolution
0.01
True RMS Voltage
Range
0.0 - 300.0
V
RMS
Accuracy
0.1 % FS + 0.05 % reading
Resolution
0.1
V
RMS
Temp
Coefficients
0.05
% FS /
°C
Real Power
Ranges
0 - 800.0
0 - 2000
W
W
Accuracy
0.2 % FS + 0.5 % reading
Resolution
0.2
1
W
W
Temp
Coefficients
0.06
% FS /
°C
Apparent Power
Available over bus only. Requires interface option
Ranges
0 - 800.0
0 - 2000
VA
VA
Accuracy
0.3 % FS + 0.8 % reading
Resolution
1
VA