2 dam, range ii offset and gain, items 11 and 12, Dam, range ii offset and gain, items 11 and 12 -9, Dam, r – AMETEK SL Series Calibration Manual User Manual
Page 17: Ii o

Elgar Electronics Corporation
Calibration
2.4.2
DAM, R
ANGE
II O
FFSET AND
G
AIN
, I
TEMS
11
AND
12
1.
LOAD OFF.
2.
Press the
© key to go to Item 11.
3.
PRES ON.
4.
Set the load current as shown in Table 2-11 for the model being
calibrated.
5.
PRES OFF.
6.
LOAD ON.
7.
Adjust using the
keys until the load DAM matches the DMM
reading.
MODEL Current
Setting
MODEL Current
Setting
SLM-60-30-150 0.00A SLH-60-120-600 0.00A
SLM-60-60-300 0.00A SLH-60-120-1200
0.00A
SLM-250-10-300 0.00A SLH-60-120-1800 0.00A
SLM-500-10-300 0.00A SLH-60-240-1200 0.00A
SLM-60-15-75 0.00A
SLH-60-240-1800
0.00A
SLH-60-360-1800
0.00A
SLH-500-60-1800
0.00A
Table 2-11. DAM, Range II Offset Adjustment Settings (Item 11)
8.
Load OFF.
9.
Press the
© key to go to Item 12.
SLM/SLH Calibration Manual
2-9