AMETEK SL Series Calibration Manual User Manual
Page 36

Calibration
Elgar Electronics Corporation
2.6.2
CV M
ODE
, L
EVEL
H
IGH
, O
FFSET AND
G
AIN
, I
TEMS
31
AND
32
Set the DC power supply current limit to 0.1A.
1.
LOAD OFF.
2.
Press the
© key to go to Item 31.
3.
PRES ON.
4.
Set voltage to the V Input value shown in Table 2-31 for the model
being calibrated.
5.
PRES OFF.
6.
LOAD ON.
7.
Adjust using the
keys until the DVM reading and the Voltage
Setting match.
V Input Voltage Setting
MODEL
MODEL
V Input Voltage Setting
SLM-60-30-150
61.5
0.800V SLH-60-120-600 61.5 0.800V
SLM-60-60-300
61.5
0.800V SLH-60-120-1200 61.5 0.800V
SLM-250-10-300
61.5
8.000V SLH-60-120-1800 61.5 0.800V
SLM-500-10-300
61.5
16.008V SLH-60-240-1200 61.5
0.800V
SLM-60-15-75
61.5
0.800V SLH-60-240-1800 61.5 0.800V
SLH-60-360-1800
61.5
0.800V
SLH-500-60-1800
61.5
16.008V
Table 2-31. CV Mode, Level High, Offset Adjustment Settings (Item 31)
8.
LOAD OFF.
9.
Press the
© key to go to Item 32.
2-28
SLM/SLH Calibration Manual