BECKHOFF KL3681 User Manual

Page 10

Advertising
background image

Product overview

Filter

The built­in filter averages the values for the last half second. If the measuring range is changed the filter is
reset. The filter is active by default. If the filter is switched off the terminal supplies approx. 16 measured
values per second.

Mains frequency

The integration times are adapted to the mains frequency (50, 60 Hz). The default is 50 Hz.

Presentation

The presentation of the values in the process data can be changed.

left aligned (0)
Left­justified presentation, full­scale value corresponds to 0x7FFFE000 (0x7FFFFFFF) or 0x80000000.

right aligned (1)
Right­aligned presentation: 18 bit resolution of the measuring range, end value corresponds to 0x0003FFFF
(262143

dec

) or 0xFFFC0000 (­262144

dec

).

In the extended measuring range these limits are exceeded!

scaled (2)
Scaled presentation: this presentation corresponds to 1 bit per µV / µA. Examples:
­ Mode 0, Autorange, 2.5 V: 2500000

dec

­ Mode 2, 800 mA: 800000

dec

float (5)
Scaled presentation in floating­point format

Scaled presentation (2) is set as default.

Note

Please note

• Only in scaled presentation the factors 0.1/1/10/100 are taken into account between the

measuring ranges.

• In left aligned presentation the extended measuring range is not available.

Zero Compenzation Interval

The compensation function compensates the internal offset drift of the ADC. The interval length is selectable.

If setting 4 "Triggered by PDO" is used, the time for reading the internal offset can be selected manually via
the "Start Calibration" process data bit.

Note

50 Hz mode

In 50 Hz mode this function should not be used (0: off).

Calibration

The terminal features manufacturer and user calibration options, which can be switched on and off
independently. The calibration always applies to all measuring ranges. Adaptation in 60 Hz mode and
compensation of the AC system offset cannot be switched off.