Modular expandability, Temp. calibration, Auto power off – Fieldpiece SC420 SC440 Essential Clamp Meters User Manual

Page 2: Auto hold, Battery replacement, Maintenance, Limited warranty, Obtaining service, Microamps dc (µadc), Frequency (hz) through leads

Advertising
background image

11

16

12

17

13

18

14

19

15

20

MicroAmps DC (µADC)

Microamps for flame rectifier diode test on a heater control.

Connect leads between flame sensor probe and control module and

turn heating unit on to read µA measurement. When the flame is on,

there should be a measurable µADC signal, typically under 10µADC.

Compare measurement to manufacturer’s specification to determine

if replacement is necessary.

Ranges: 500µA Resolution: 0.1µA

Accuracy: ±(1.0% + 2) Voltage burden: 1V

Overload Protection: 600VDC or 600VAC rms

Frequency (Hz) Through Leads

Check incoming voltages to ensure they are cycling at 60Hz. For

frequency measurements on VFD equipment, use the amp clamp.

Ranges: 500Hz, 5kHz, 50kHz, 500kHz, 1MHz Resolution: 0.1Hz

Accuracy: ±(0.1% + 5) Sensitivity: 10Hz to 1MHz: >3.5Vrms

PW: >1µs Duty Cycle Limits: >30% and <70%

Duty Cycle (%)

Duty cycle shows the % On Time of a 5V logic signal square wave.

Ranges: 5%-95% (40Hz to 10kHz), 10%-90% (10kHz to 20kHz)

Accuracy (5V logic): ±(2% + 10) Resolution: 0.1% PW: >10µs

Overload Protection: 600VDC or 600VAC rms

Capacitance (MFD)

Set to MFD to test motor start and run capacitors. Capacitors

are one of the most failure prone components in a HVACR system.

Disconnect from power and resistors between terminals. Discharge

capacitor before testing. If dIS.C is displayed, capacitor is not

discharged completely.

Ranges: 5µF, 50µF, 500µF, 5mF Resolution: 1nF

Accuracy: ±(3% + 15) 5µF, ±(3% + 5) 50µF to 500µF, ±(5% + 20) 5mF

Overload Protection: 600VDC or 600VAC rms

Voltage DC (VDC)

Select VDC and measure DC voltages on circuit boards on more

advanced HVACR systems.

Ranges: 500mV, 5V, 50V, 500V, 600V Resolution: 0.1mV

Accuracy: ±(0.5% + 2)

Input impedance: >100MΩ (500mV), 10MΩ (5V),

9.1MΩ (50V-600V)

Resistance (Ω)

Used for “ohming out” a compressor. 0.1Ω resolution is necessary

to test the resistance between the motor poles because the values are

typically very low.

Ranges: 500Ω, 5kΩ, 50kΩ, 500kΩ, 5MΩ, 50MΩ

Resolution: 0.1Ω Overload Protection: 600VDC or 600VAC rms

Accuracy: ±(1.0% + 5) 500Ω to 500kΩ, ±(1.5% + 5) 5MΩ, ±(3.0%

+ 5) 50MΩ

Continuity (

O

F

F

MFD

NCV

V

AC

V

DC

A

AC

Hz

Hz

%

°C

°F

SC56

AUTO OFF

TRUE RMS

CAT III

600V

TEMP

K-TYPE

30V

MAX

SEL

HOLD

INRUSH

)

Use the continuity feature to test if a circuit is open or closed. Use

this feature to check fuses as well. A steady “beep” and green LED

indicate you have continuity. Press SELECT once.

Range: 500Ω Resolution: 0.1Ω Response time: 100ms

Audible beep: <30Ω Overload Protection: 600VDC or 600VAC rms

Diode Test (

O

F

F

MFD

NCV

V

AC

V

DC

A

AC

Hz

Hz

%

°C

°F

SC56

AUTO OFF

TRUE RMS

CAT III

600V

TEMP

K-TYPE

30V

MAX

SEL

HOLD

INRUSH

)

Test diodes for proper forward and reversed-biased functions.

Test current: 0.8mA (Approx.) Accuracy: ±(1.5% + 5)

Open circuit volts: 3.2VDC typical Audible beep: <0.03V

Visual Indicator: Green LED

Overload Protection: 600VDC or 600VAC rms

Amps AC (AAC) True RMS

Test any isolated power line. Press SELECT on VAC/AAC/Hz

position. Read AAC in upper display. True RMS on SC440 only.

Ranges: 50A, 400A Resolution: 0.01A Crest factor: ≤ 3

Accuracy: ±(2.0% + 10) 50-60Hz Jaw Opening: 1.2in (30 mm)

Frequency (Hz) Through Clamp

Measure Hz on variable frequency drive motors. Turn dial to VAC/

AAC/Hz and press SELECT twice. Hz will show in upper display.

Range: 10Hz to 400Hz Accuracy: ±(0.1% + 5)

Minimum current range: > 5AAC Overload Protection: 400AAC

Resolution: 0.1Hz

Inrush Current (SC440 only)

Inrush feature captures starting current of a compressor motor.

Starting current can assist in diagnosing a motor before it fails.

To activate Inrush feature

1. Turn selector switch to VAC/AAC/Hz

2. Press SELECT once to show AAC on upper display.

3. Press INRUSH once to select 50AAC range. Press INRUSH twice to

select 400AAC range.

4. Clamp meter around compressor start wire. Turn motor on. The

starting current will hold on the upper display.

5. Hold INRUSH for 2 seconds to exit Inrush feature or press and

release INRUSH to measure starting current again.

Inrush measurement period: 100-milliseconds

Minimum input: >2A on 50A range; >20A on 400A range

!

Note: AAC through the clamp and voltage through the test

leads can be measured simultaneously. However, if only AAC,

Frequency (Hz), or Inrush is measured through the clamp, test

leads and thermcouple must be unplugged from the meter.

Modular

Expandability

Your new Swivel Clamp Meter is compatible

with all Fieldpiece Accessory Heads. With

Fieldpiece Accessory Heads, you can measure

any available parameter, and read the

measurement on your new meter's display in

real-time.

Just set the range to VDC and press the

RANGE button until mV is displayed. Remove

the probe tips of your test leads, and connect

your accessory head (model AAV3 shown).

Visit www.fieldpiece.com to see all of the

different accessory heads that Fieldpiece offers.

Temp. Calibration

For accuracies of ±1°F, calibrate to a known

temperature. A glass of stabilized ice water

is very close to 32°F (0°C) and is usually very

convenient but any known temperature can

be used.

1. Select the °F °C range.

2. Plug thermocouple to be

calibrated into the Type K jack.

3. Unscrew A and B and remove

the battery cover.

4. Stabilize a large cup of ice

water. Stir the ice with the

water until temperature stays

at a stable value.

5. Immerse the thermocouple

probe and let it stabilize.

Keep stirring water to prevent

micro-environments.

6. Do not let the thermocouple

come in direct contact with

ice.

7. Use a small screwdriver to

adjust calibration pot C below

the battery as close to 32°F

(0°C) as you would like.

Note: J1-J2 switch is for auto-

calibration purposes only. Do

not switch from J2.

Auto Power Off

Auto power off or APO will automatcially turn

off your meter after 30 minutes of inactivty. By

default it is activated and APO will show on the

display. To disable, turn meter off. Hold

HOL

D

COM

CAT III

600V

Fieldpiece

Select

Range

M/m

SC240

V

AC

NCV

A

AC

°F°C

V

DC

MFD

TEMP

AUTO OFF

TYPE K

30V

MAX

and

power on the meter by turning the selector dial

to any range. Release

HOL

D

COM

CAT III

600V

Fieldpiece

Select

Range

M/m

SC240

V

AC

NCV

A

AC

°F°C

V

DC

MFD

TEMP

AUTO OFF

TYPE K

30V

MAX

after the beep. APO will

no longer display over the battery icon.

Auto Hold

Press and hold

OFF

V

AC

V

AC

%

NCV

V

DC

µA

DC

MFD

H

Z

A

AC

50

H

Z

A

AC

400

°F

°C

H

Z

SC640

AUTO OFF

TYPE K

30V

MAX

Select

Range

Hold

INRUSH

COM

TRUE RMS

for two seconds. Meter

will beep and HOLD will blink on LCD. After

6 seconds the measurement on screen will

freeze automatically. Press

OFF

V

AC

V

AC

%

NCV

V

DC

µA

DC

MFD

H

Z

A

AC

50

H

Z

A

AC

400

°F

°C

H

Z

SC640

AUTO OFF

TYPE K

30V

MAX

Select

Range

Hold

INRUSH

COM

TRUE RMS

to exit this mode.

Battery Replacement

When your meter’s battery is low, the battery

icon will appear empty and blink for 30 seconds.

“LO.bt” will display and meter will power off.

Turn dial to OFF position, disconnect test

leads and remove the battery cover with magnet

strap (SC440 only) on the back of your meter.

Remove old battery and replace with a standard

9V battery only. Be sure to re-insert the magnet

strip before re-installing the battery cover.

Maintenance

Clean the exterior with a dry cloth. Do not

use liquid.

Limited Warranty

This meter is warranted against defects in

material or workmanship for one year from

date of purchase. Fieldpiece will replace or

repair the defective unit, at its option, subject

to verification of the defect.

This warranty does not apply to defects

resulting from abuse, neglect, accident,

unauthorized repair, alteration, or unreasonable

use of the instrument.

Any implied warranties arising from the sale of

a Fieldpiece product, including but not limited

to implied warranties of merchantability and

fitness for a particular purpose, are limited to

the above. Fieldpiece shall not be liable for loss

of use of the instrument or other incidental or

consequential damages, expenses, or economic

loss, or for any claim of such damage, expenses,

or economic loss.

State laws vary. The above limitations or

exclusions may not apply to you.

Obtaining Service

Call Fieldpiece Instruments for one-price-

fix-all out-of-warranty service pricing. Send

check or money order for the amount quoted.

Send the meter freight prepaid to Fieldpiece

Instruments. Send proof of date and location

of purchase for in-warranty service. The meter

will be repaired or replaced, at the option

of Fieldpiece, and returned via least cost

transportation.

For international customers, warranty for

products purchased outside of the U.S. should

be handled through local distributors.

www.fieldpiece.com

© Fieldpiece Instruments, Inc 2014; v21

V

AC

NCV

MFD

H

Z

A

AC

°F°C

%

V

DC

µA

DC

H

Z

V

AC

NCV

MFD

H

Z

A

AC

°F°C

%

V

DC

µA

DC

H

Z

V

AC

NCV

MFD

H

Z

A

AC

°F°C

%

V

DC

µA

DC

H

Z

V

AC

NCV

MFD

H

Z

A

AC

°F°C

%

V

DC

µA

DC

H

Z

V

AC

NCV

MFD

H

Z

A

AC

°F°C

%

V

DC

µA

DC

H

Z

O

FF

V

A

C

V

A

C

%

N

C

V

V

D

C

µA

D

C

M

FD

H

Z

A

A

C

50

H

Z

A

A

C

40

0

°F

°C

H

Z

SC6

4

0

AU

TO

O

FF

TY

PE

K

30

V

M

AX

Se

le

ct

Ra

ng

e

Ho

ld

INRU

SH

C

O

M

TR

UE

R

M

S

V

AC

NCV

MFD

H

Z

A

AC

°F°C

%

V

DC

µA

DC

H

Z

Discharge Cap First!

SC440

Fieldpiec

e

TRUE RMS

VAC

NCV

MFD

HZ

AAC

°F°C

Select

Range

Hold

%

VDC

µADC

HZ

TEMP

AUTO OFF

COM

TYPE K

30V

MAX

CAT III

600V

V

AC

NCV

MFD

H

Z

A

AC

°F°C

%

V

DC

µA

DC

H

Z

Air Velocity

& Temperature

Head

English

READ

LO BATT

ON

LCD X

100

Average

(16 sec)

Metric

Real

time

AAV3

AUTO-

OFF

English

Metric

Ft/min

M/s

KM/hr

MPH

ºF

ºC

OFF

TEMP CAL

DO NOT SWITCH

FROM J2

A

B

C

V

AC

NCV

MFD

H

Z

A

AC

°F°C

%

V

DC

µA

DC

H

Z

V

AC

NCV

MFD

H

Z

A

AC

°F°C

%

V

DC

µA

DC

H

Z

Advertising
This manual is related to the following products: