Fieldpiece SPDM1 - Pocket Digital Multimeter User Manual

Maintenance, Limited warranty, Service

Advertising
background image

AUTORANGING POCKET

DIGITAL MULTIMETER: SPDM1

OPERATOR’S MANUAL

For your safety...

General: Inspect the test leads for damage to the

insulation or exposed metal. Do not use if the meter
or test leads look damaged, or if you suspect that the
meter is not operating properly. Never ground your-
self when taking electrical measurements. Do not
touch exposed metal pipes, outlets, fixtures, etc.,
which might be at ground potential. Keep your body
isolated from ground by using dry clothing, rubber
shoes, rubber mats, or any approved insulating mate-
rial. When disconnecting from a circuit, disconnect
the "RED" lead first, then the common lead. Work
with others. Turn off power to the circuit under test
before cutting, unsoldering, or breaking the circuit.
Keep your fingers behind the finger guards on the
probes. Do not measure resistance when circuit is
powered. Do not apply more than rated voltage
between input and ground.

All voltage tests: All voltage ranges will withstand

up to 600V. Do not apply more than 600VDC or
600VAC.

AC tests: Disconnect the meter from the circuit

before turning any inductor off, including motors,
transformers, and solenoids. High voltage transients
can damage the meter beyond repair. Do not use
during electrical storms.

Maintenance

Clean the exterior with clean dry cloth. Do not use

liquid.

Battery replacement: When the multimeter dis-

plays " " the battery must be replaced. Disconnect
leads from test points, turn meter off, and remove the
battery cover. Replace the battery with 2 NEDA #
1166A 1.5V batteries.

Limited warranty

This meter is warranted against defects in materi-

al or workmanship for one year from date of pur-
chase. Fieldpiece will replace or repair the defective
unit, at its option, subject to verification of the defect.

This warranty does not apply to defects resulting

from abuse, neglect, accident, unauthorized repair,
alteration, or unreasonable use of the instrument.

Any implied warranties arising from the sale of a

Fieldpiece product, including but not limited to
implied warranties of merchantability and fitness for a
particular purpose, are limited to the above.
Fieldpiece shall not be liable for loss of use of the
instrument or other incidental or consequential dam-
ages, expenses, or economic loss, or for any claim of
such damage, expenses, or economic loss.

Laws vary with jurisdiction. The above limitations

or exclusions may not apply to you.

Service

Return any defective SPDM1 Series meter to

Fieldpiece for warranty service along with proof of
purchase. Contact Fieldpiece for out of warranty
repair charges.

www.fieldpiece.com

Symbols used:

Caution, refer to manual.

Ground

Double insulation

Alternating current

Direct current

SPECIFICATIONS

Display: 3½ digit liquid crystal display (LCD) with a

maximum reading of 1999.

Polarity: Automatic, (-) polarity indication.
Overrange: "OL" mark indication.
Auto power off: approx. 10 minutes.
Operating environment:32

°F to 104°F at <70%RH.

Storage temperature: -4

°F to 140°F, 0 to 80% R.H.

with battery removed.

Temperature coefficient: 0.1

×(specified accuracy)

per

°F. (32°F to 64°F, 82°F to 104°F).

Altitude: 6561.7 feet (2000m)
Power: Two 1.5V button-type batteries (IEC # LR-

44, EDA # 1166A).

Battery life: 70 hours continuous operation.
Measurement rate: 2 times per second, nominal.
Safety:UL61010B-1, UL61010B-2-031, EN61010-1,

EN61010-2-031, CAT II 600V, CAT III 300V, Class
2, Pollution degree II, Indoor use, CE, C-Tick.

Resistance (autoranging)

Ranges: 200

Ω, 2kΩ, 20kΩ, 200kΩ, 2MΩ, 20MΩ

Resolution: 100m

Ω

Accuracy:

±(2.0% rdg + 5 dgts) on 200Ω range

±(2.0% rdg + 4 dgts) on 2kΩ, 20kΩ, 200kΩ ranges

±(3.0% rdg + 4 dgts) on 2MΩ range

±(5.0% rdg + 5 dgts) on 20MΩ range

Open circuit volts: -0.45VDC typical, (-1.2VDC on

200

Ω range)

Overload protection: 450VDC or AC rms

Continuity

Audible indication: Less than 25

Ω

Response time: 500ms
Overload protection: 450VDC or AC rms

Diode test

Audible indication: Less than .25V
Range: 2V
Resolution: 10mV
Accuracy:

±(3.0% rdg + 3 dgts)

Test current: 1.2mA
Overload protection: 450VDC or AC rms

DC volts (autoranging)

Ranges: 2V, 20V, 200V, 600V
Resolution: 1mV
Accuracy:

±(2.0% rdg + 2 dgt)

Input impedance: 10M

Ω on 2V, 9.1MΩ all other

Overload protection: 600VDC or AC rms

AC volts 50/60Hz (autoranging)

Ranges: 2V, 20V, 200V, 600V
Resolution: 1mV
Accuracy:

±(4.0% rdg + 5 dgts)

Input impedance: 10M

Ω on 2V, 9.1MΩ all other

Overload protection: 600V DC or AC rms

WARNINGS

DISCONNECT TEST LEADS FROM CIRCUIT

before attempting to replace the battery.

TEST NCV FUNCTION ON KNOWN LIVE

WIRE before using.

!

!

!

© Fieldpiece Instruments, Inc. 2010 v09

Advertising