Fieldpiece SRH3 - In Duct Diagnostic Psychrometer User Manual

Fieldpiece, In-duct hot wire anemometer, In-duct problem-solving psychrometer

Advertising
background image

01

03

06

02

05

04

07

09

12

08

11

10

Quick Start

1. Power on your SRH3 by holding the

ON/OFF button for 1 second.

2. Use the TSH/TEET/NORM button to

select NORMAL mode.

3. Press the RH/DP/WB button to cycle

between relative humidity, dew

point, and wet bulb temperatures.

4. The air temperature will always be

shown in the lower display and the

relative humidity, dew point, or wet

bulb measurement will be displayed

in the upper display in real-time.

5. Insert sensor probe into duct for

in-duct measurements.

Certifications

C-Tick (N22675)

CE

WEEE

RoHS Compliant

Description

With your purchase of the SRH3 you

now have the ability to take supply/

return temperature, %RH, wet bulb

and dew point measurements from

inside the duct or plenum.

Your SRH3 is a portable, hand held,

diagnostic psychrometer, designed

for the HVAC/R technician.

SRH3 helps you properly charge

a fixed restrictor system by quickly

determining the Target Superheat

(TSH). Your SRH3 can also help

determine optimum airflow across

the evaporator by easily performing

a Target Evaporator Exit Temperature

(TEET) test.

The 38"(96cm) telescoping probe

with laser etched ruling and flattened

edges allows you to locate proper

measurement points within a duct and

ensures that your probe is properly

aligned.

The dual display with a bright

backlight, rugged rubber boot with

probe clips, make sure the SRH3 is

ready for any job.

Display

Auto Power Off Enabled

Low Battery Indicator

Relative Humidity

Dew Point

Wet Bulb

Dry Bulb

%Relative Humidity

Outdoor

Indoor

Target Superheat

Target Evaporator Exit Tem-

perature

Normal Range

80

STA2

UNITS

CLEAR DATA

RECORD MODE

PRESS FOR

1 SECOND

In-Duct

Hot Wire

Anemometer

DUCT

ON/OFF

ENTER

HOLD

AVERAGE

MAX/MIN

AUTO-OFF

Temperature (Fahrenheit)

80

STA2

UNITS

CLEAR DATA

RECORD MODE

PRESS FOR

1 SECOND

In-Duct

Hot Wire

Anemometer

DUCT

ON/OFF

ENTER

HOLD

AVERAGE

MAX/MIN

AUTO-OFF

Temperature (Celsius)

Hold Display

Maximum Display

Minimum Display

Fieldpiece

In-Duct

Problem-Solving

Psychrometer

OPERATOR'S

MANUAL

Model SRH3

Controls

Hold 1 second to toggle power on/off.

Toggle Backlight. (Hold while powering

your SRH3 on to disable APO.)

Toggle display to show relative

humidiy, dew point and wet

bulb temperatures

Toggle between Fahrenheit and Celcius.

Hold for 2 seconds to display battery

percentage.

Toggles to display Maximum, Minimum,

or Hold the current reading. Hold for 1

second to exit and clear stored values.

Confirm Selection

Outputs calculations for Target

Superheat or Target Evaporator Exit

Temperature

Toggle to enter Target Superheat,

Target Evaporator Exit Temp, or

Normal modes.

Functions

Backlight

1. Press the backlight button to toggle the backlight

on/off at any time during the SRH3's use.

Note: The back light will automatically shut off after 1

minute to conserve battery life.

Maximum/Minimum/Hold

1. Pressing the MAX/MIN/H button activates the

Max-Min-Hold function, holding the maximum and

minimum measured values until cleared.

2. Once the Max-Min-Hold function has been

enabled, pressing MAX/MIN/H cycles between

displaying maximum (MAX), minimum (MIN) and

freezing the current value.

3. Press and hold the MAX/MIN/H button for 2 seconds

to erases all stored maximum and minimum values

and return to real-time measurements.

Units

1. Press the units button to switch between degrees

Fahrenheit and Celsius.

Battery Check

1. Press and hold the UNIT button for 2 seconds to

display the battery life in percentage.

Output

1. Press the OUTPUT button to perform calculations

of Target Superheat and Target Evaporator Exit

Temperature after necessary measurements have

been entered.

How to Use

Normal Mode

Normal mode allows you quickly

and easily take real-time dry bulb, wet

bulb, dew point, and relative humidity

measurements.

1. Press TSH/TEET/NORM button until NORM is

displayed in lower left corner.

2. For upper display measurements, press the RH/DP/

WB button to toggle relative humidity, dew point,

or wet bulb temperature measurements.

Note: Dry bulb (air temperature) will always show in

the lower display.

Target Superheat Mode

Perform a simple Target Superheat

test to determine what the superheat

should be on a fixed restrictor system.

Use the target superheat to compare

to the actual superheat to determine

if a system is properly charged.

1. Press TSH/TEET/NORM button until TSH is displayed

in the lower left of the LCD.

2. Take a wet bulb measurement at the inlet of the

evaporator. Generally right before the filter is a

good location. Drill or punch a 3/8" hole into the

return plenum for a measurment depth of up to

24 inches. See Figure 1 for recommended return

plenum placement of probe. Insert SRH3 probe into

the plenum to take the wet bulb measurement.

Press ENTER to lock in the reading.

3. Take a dry bulb (air temperature) measurement

going into the condensor. Press ENTER to lock in

the reading.

4. Press OUTPUT to calculate the target superheat.

Compare the target superheat to the actual

superheat of the system.

5. Seal any holes before leaving the jobsite.

Target Evaporator Exit

Temperature Mode

Target Evaporator Exit Temperature

(TEET), aka temperature drop or

delta T, is used to determine if the

evaporator is getting the optimum

airflow. A TEET test can be quickly

performed by taking two easy

temperature measurements at the

return side of the evaporator.

1. Press the TSH/TEET/NORM button until TEET is

displayed in the lower left of the LCD.

2. Take a wet bulb measurement inside the return

plenum. Drill or punch a 3/8" hole into the return

plenum. See Figure 1 for recommended return

plenum placement of probe. Insert SRH3 probe into

the plenum to take the wet bulb measurement.

Press ENTER to lock in the reading.

3. Take a dry bulb measurement at the same place the

previous wet bulb measurement was taken. Press

ENTER to lock in the reading.

4. Press OUTPUT to calculate the target evaporator exit

temperature. Compare this calculated target to the

actual evaporator exit temperature.

5. Take a dry bulb measurement inside the supply

plenum to check the actual Evaporator Exit

Temperature. See Figure 1 for recommended supply

plenum placement of probe. Drill or punch a 3/8"

hole into the supply plenum. Press TSH/TEET/NORM

until NORM is displayed.Insert SRH3 probe into

the plenum and read the temperature in the lower

display.

6. Seal any holes before leaving the jobsite.

Note: For both Target Superheat and Target Evaporator

Exit Temperature modes, you can adjust any

temperature measurement after calculations have

been performed by pressing ENTER. This prompts

you to re-take the measurements.

RCONE1 Probe Lock

Use the RCONE1 to lock your sensor

probe in place inside the duct. Using

the RCONE1 and magnet on the SRH3

allows you hands-free in-duct testing.

See Figure 2.

Specifications

Telescoping Probe Length: Up to 38 inches (97cm)

Probe Tip Diameter: 0.35 inch (9mm)

Storage temperature: -4°F to 140°F (-20°C to 60°C), 0

to 80% RH (with battery removed)

Temperature Coefficient: 0.1 x (specified accuracy)/°C

(<18°C or >28°C)

Over range: "OL" or "-OL" is displayed

Power: Single standard 9-volt battery, NEDA 1604, JIS

006P, IEC 6F22

Auto Power off: after 15 minutes of inactivity if APO

is active.

Battery life: 150 hours typical (alkaline)

Low Battery Indication: is displayed when the

battery voltage drops below the operating level.

Dimensions: 7.9 in (H) x 2.6 in (W) x 1.4 in (D),

[200mm (H) x 66mm (W) x 36mm (D)]

Weight: Approx. 400g, including battery

Temperature:

Sensor type: Precision thermistor

Operating environment: -4°F to 140°F (-20°C to 60°C)

Range: -4°F to 140°F (-20°C to 60°C)

Resolution: 0.1°F / 0.1°C

Accuracy: ±(1°F) 32°F to 113°F

±(2°F) -4°F to 32°F, 113°F to 140°F

±(0.5°C) 0°C to 45°C

±(1°C) -20°C to 0°C, 45°C to 60°C

Relative Humidity:

Sensor Type: Capacitance polymer film

Operating environment: 32°F to 131°F (0°C to 55°C)

Range: 0% to 100%RH

Accuracy: ±(2.5%) 10% to 90%RH

±(5%) <10%RH and >90%RH

Note: Above accuracies stated at 73.4°F (23°C).

Sensor Response Time: 60 seconds typical for 90% of

total range.

Sensor Hystersis: ±1%RH typical (Excursion of 10% to

90% to 10%RH)

80

SRH3

PRESS 1 SEC

TO CLEAR

PRESS FOR

1 SECOND

In-Duct

Diagnostic

Psychrometer

UNITS

ON/OFF

TSH/TEET/NORM

RH/DP/WB

AUTO-OFF

BATTERY

CHECK

ENTER

MAX/MIN/H

OUTPUT

80

80

STA2

UNITS

CLEAR DATA

RECORD MODE

PRESS FOR

1 SECOND

In-Duct

Hot Wire

Anemometer

DUCT

ON/OFF

ENTER

HOLD

AVERAGE

MAX/MIN

AUTO-OFF

80

STA2

UNITS

CLEAR DATA

RECORD MODE

PRESS FOR

1 SECOND

In-Duct

Hot Wire

Anemometer

DUCT

ON/OFF

ENTER

HOLD

AVERAGE

MAX/MIN

AUTO-OFF

80

STA2

UNITS

CLEAR DATA

RECORD MODE

PRESS FOR

1 SECOND

In-Duct

Hot Wire

Anemometer

DUCT

ON/OFF

ENTER

HOLD

AVERAGE

MAX/MIN

AUTO-OFF

80

STA2

UNITS

CLEAR DATA

RECORD MODE

PRESS FOR

1 SECOND

In-Duct

Hot Wire

Anemometer

DUCT

ON/OFF

ENTER

HOLD

AVERAGE

MAX/MIN

AUTO-OFF

RH/DP/WB

UNITS

BATTERY

CHECK

PRESS 1 SEC

TO CLEAR

MAX/MIN/H

ENTER

OUTPUT

TSH/TEET/NORM

ON/OFF

Recommended Hole

Placement for Supply

Plenum Measurements

Recommended

Hole Placement for

Return Plenum

Measurement is

directly before the

evaporator coil. Look behind

the panel before drilling and make

sure you do NOT drill a hole through any critical HVAC compo-

nents.

Figure 1

SRH3

PRESS 1 SEC

TO CLEAR

PRESS FOR

1 SECOND

In-Duct

Diagnostic

Psychrometer

UNITS

ON/OFF

TSH/TEET/NORM

RH/DP/WB

AUTO-OFF

BATTERY

CHECK

ENTER

MAX/MIN/H

OUTPUT

ARH5

In-Duct

Psychrometer

Head

ON

AUTO OFF

1%RH/mVDC

1°F/mVDC

1°C/mVDC

LO BATT

DEW POINT

%RH

WET BULB

TEMP

°F

°C

ON

OFF

ET2W

LO BATT

SEND

RECEIVE

ON

AC

DC

Wireless

Transmitter

SYNC

Clear

Input

Output

Recall

Enter

Save

Sync

SETUP

Target

Evaporator

Logger

Light

Exit Temp

CFM

Display

Superheat

Subcooling

Combustion

HVAC Guide

CheckMe!

Fieldpiece

HG3

Data

Service

E

Exit T

Ex

Ex

E

e

Target

arget

arget

rget

rget

get

Targe

or

ator

ato

ator

ator

ator

Ta

Evaporat

t Temp

INPUT FORM

SH Table: Standard

Refrigerant: R-22

OD Dry Bulb:

85.4°F

Return WB:

63.0°F

SL Pressure:

52.4psig

SL Temp:

45.4°F

CustomerID:

__JONES123

Use the RCONE1 for

hands-free in-duct

measuremesnts.

Figure 2

WARNING

Do not retract the sensor probe by pulling on the cord.

Doing so may sever the cord from the sensors.

Advertising