Temperature measurement, Barometric pressure – Fluke Biomedical DPM4 User Manual

Page 13

Advertising
background image

Parameter Tester

Specifications

5

Resolution ...........................................................0.5 mmHg
Units of measure .................................................mmHg, mBar, cmH2O, PSI, InHg, InH2O, kgcm2, and kPa

Temperature Measurement

Operating range.....................................................-40 °C to +200 °C
Accuracy ................................................................±(2 % of reading, +0.5 °C)
Resolution..............................................................0.1 °C and °F
Units of measure ...................................................°C and °F
Use standard external temperature probe type PT-100 or PT-1000 (DIN/IEC 751 Class A) for temperature measurements in °C or °F.
PT-100 Temperature Probe

Operating range ..................................................-50 °C to +200 °C
Accuracy..............................................................0.13 °C @ -100 °C, 0.1 °C @ 0 °C, 0.2 °C @ 100 °C

PT-1000 Temperature Probe

Operating range ..................................................-50 °C to +200 °C
Accuracy..............................................................0.3 °C

Barometric Pressure

DPM4 Model 2G or 2H
It is possible to compensate for sea level and calibrate for offsets.

Operating range ..................................................380 to 900 mmHg
Accuracy..............................................................2 % of reading ±1 LSD
Resolution ...........................................................1 mmHg
Units of measure .................................................mmHg, mBar, inHg, and hPa