15

Table 3. Gas Flow and Temperature Tests

Number Description

A

Gas (F1): Sets the type of gas you are using for the measurement.

Note

This parameter must be set correctly to make accurate measurements.

Unit (F2): Sets SCCM, ml/min for gas flow, and °C/F for the temperature.
Peak (F3): The Tester displays Peaktest as long as the measurement is continuing. The Tester displays the
highest measurement value. Reset the measurement with the F4 key.
Zero (F4): Resets the flow measurement. This function should only be used to set zero, and not for relative
measurements in proportion to a given pressure.

B
C

Press

< MENU SCROLL or MENU SCROLL > to view the available types of gas.

Select the required type of gas by pressing the appropriate F1-F4 key.
Press

< MENU SCROLL to validate the selection.

D

Select the unit of gas flow with the F1 or F2 key, and the unit of temperature with the F4 key.
Press

< MENU SCROLL to confirm the selection(s).

Note

An optional PT-100 or PT-1000 temperature probe is necessary for the temperature measurement.

E

Press

< MENU SCROLL to return to the preceding screen.

Reset resets the measurement in progress to zero

F

Zeros the flow measurement.