Gas flow, Relative humidity, Controls and connections – Fluke Biomedical DPM4 User Manual

Page 14

Advertising
background image

DPM4
Users Manual

6

Gas Flow

DPM4 Model 2G or 2H
These models measure with an embedded sensor with 11 calibration points to compensate for nonlinearity. The Tester stores calibration
constants in firmware.

Operating range ..................................................-750 to +750 ml/min
Accuracy .............................................................1 % of range or 5 % of reading, whichever is greater
Resolution ...........................................................0.1 ml/min
Compatibility........................................................Air, N2, O2, CO2, H2, and He
Units of measure .................................................ml/min (or SCCM - Standard Cubic Centimeters per Minute)

Relative Humidity

DPM4 Model 2G or 2H
An integrated sensor in the Tester determines relative humidity measurements.

Operating range ..................................................12 to 95 % RH
Accuracy .............................................................3.5 % of reading ±2 % RH @ 25 °C for RH between 12 % and 95 % RH
Resolution ...........................................................1 % RH
Gas compatibility.................................................Air
Units of measure .................................................% RH

Controls and Connections

This section describes the controls and connections of
the Tester. Refer to Figure 1 and Table 1.