Infusion Device Analyzer

Safety Information

3

Disable the Product if it is damaged.
Remove the batteries to disable the
Product.

Do not use the Product if it is damaged.

Do not use the Product on infusion
devices that are attached to patients.

Do not reuse test tubing or syringes for
patient infusion.

Avoid possible contamination of
reusable components due to backflow
conditions. Some older style infusion
devices may have reusable components
that could come in direct contact with
the fluids being pumped. When testing
these types of devices take care to avoid
possible contamination of reusable
components.

Caution

To prevent possible damage to the product
or to equipment under test:

Only use degassed de-ionized water with
the Product. Wetting agent may be
added.

Remove internal water before shipping
or storing. Do not use compressed air to
clean out the Product.

Do not expose the Product to
temperature extremes. For proper
operation, ambient temperatures should
be from 15

°C to 30 °C (59 °F to 86 °F).

Performance may be adversely affected
if temperatures fluctuate above or below
this range. For Storage Temperature
limits, see the Specifications section.

Do not use the Product in close
proximity to sources of strong
electromagnetic radiation (for example,
unshielded intentional RF sources).
These sources may interfere with proper
operation.