When you connect to the inlet circuits (for
example, when you attach the priming syringes
to the stopcocks) make sure no new bubbles are
introduced.

Caution

To prevent possible damage to the product
or to equipment under test , do not use
delivery set or components that have been
used for prior testing for patient infusion.

Note

Before you use the delivery set (the tubing, the
syringe, and stopcocks), make sure it is within
the specified use period of the manufacturer.
Many sets are made to be used only once.

Connect Drains

Follow these recommendations when you connect the
drain tubes to the Product outlets:

Do not allow the drain tubes to rise more than
10 cm (4 in) at any point above the height of the
inlet ports of the Product.

The discharge end of the drain tubes must not
be more than 10 cm (4 in) below the bottom of
the Product.

Product Maintenance

Warning

To prevent possible electrical shock, fire, or
personal injury:

Batteries contain hazardous chemicals
that can cause burns or explode. If
exposure to chemicals occurs, clean
with water and get medical aid.

Do not disassemble the battery.

Do not disassemble or crush battery
cells and battery packs.

Do not put battery cells and battery
packs near heat or fire. Do not put in
sunlight.

Do not short the battery terminals
together.

Do not keep cells or batteries in a
container where the terminals can be
shorted.