Performance Specifications

Average Flow Rate Measurement

Technique .............. Flow is calculated by measuring

volume over time

Range .................... 0.5 ml/h to 1000 ml/h
Accuracy ................ 1 % of reading ±1 LSD for flows of

16 ml/h to 200 ml/h for volumes
over 20 ml; otherwise 2 % of
reading ±1 LSD for volumes over
10 ml under laboratory conditions

Max test duration .... 10 hours on battery

Volume Measurement

Technique .............. Volume is measured directly by the

measuring module in minimum
sample sizes of 60 µl

Range .................... 0.06 ml to 999 ml
Accuracy ................ 1 % of reading ±1 LSD for flow

rates of 16 ml/h to 200 ml/h for
volumes over 20 ml. Otherwise
2 % of reading ±1 LSD for volumes
over 10 ml under laboratory
conditions

Max test duration .... 10 hours on battery

Pressure Measurement

Technique
(Occlusion test) ...... Direct measurement of pressure at

the inlet port

Range ..................... 0 psi to 45 psi and equivalent in

mmHg, Bar and kPa

Accuracy ................ 1 % of Full Scale ±1 LSD under

laboratory conditions

Max test duration .... 30 minutes

For a complete list of specifications see the IDA-1S Users
Manual.