Clean the product (outside), Clean the product (inside) – Fluke Biomedical IDA-1S Getting Started User Manual

Page 17

Advertising
background image

Infusion Device Analyzer

Product Maintenance

9

Remove the input signals before you
clean the Product.

Use only specified replacement parts.

Have an approved technician repair the
Product.

For safe operation and maintenance of the
product:

Repair the Product before use if the
battery leaks.

Be sure that the battery polarity is
correct to prevent battery leakage.

Use only Fluke approved power adapters
to charge the battery.

After troubleshooting or maintenance, restart the Product
and ensure it starts without errors.

Clean the Product (Outside)

To clean the outside of the Product, disconnect from the
power supply and use only a damp cloth with mild
detergent.

Clean the Product (Inside)

It is possible that microbial growth can become present in
the measuring module of the Product. It is recommended
that you clean the fluid paths at 3 month intervals. To
clean the inside of the Product, inject 20 ml of a warm
water and detergent solution into the Fluid Inlet Port. After
5 minutes, flush with clean water. Always pass water from
the fluid inlets to the outlets.