Victoreen 451 EXL
Operators Manual

3-6

time remaining in the 30-second reset period. During this period, any radiation detected by the 451 will
not be included in the dose measurement.








Integrated Dose Status Message Box During Reset

Step 4 - After the 30 second reset period, the 451 Assistant will display a status message box indicating
that the 451 is integrating. If timer controlled integration has been selected, the 451 Assistant will indicate
the elapsed integration time and the integration time remaining. If timer controlled integration has not
been selected, the 451 Assistant will only indicate the elapsed integration time. Stop can be pressed at
any time to stop the integration process and read the integrated dose measurement.










Integrated Dose Status Message Box During Integration

Step 5 After the specified timer duration or stop is pressed, the Assistant will query the instrument for the
integrated dose.










Integrated Dose Status Message Box, Integration Complete

After integration is complete, the integrated dose measurement and the average dose rate will be inserted
at the active worksheet cell in Excel as shown below.