K-edge mammo filter pack – Fluke Biomedical 199XRAY User Manual

Page 13

Advertising
background image

Using the X-RAY kVp Function

Selecting a Filter Pack

2

2-5

k-Edge Mammo Filter Pack

The k-Edge Mammo filter pack is used to determine the
correction factor (recalibrate) for the Linear Mammo filter
pack.

Proceed as follows to do the recalibration:

1

Select ■ k-Edge Mammo… with
the arrow keys, then accept.

The screen will show the present
k-Edge calibration values.

2

Accept ■ Recalibrate… to start
the calibration.

You will see the calibration startup
screen as shown below.

3

Follow the screen instructions.

4

Close the instruction banner and
start the calibration.

You will be guided through the calibration procedure by
screen messages. See also ‘Cadmium K-edge Mammo
Filter Pack Calibration Overview’ on page 2-3