Using the X-RAY kVp Function
Selecting a Filter Pack
2
2-5
k-Edge Mammo Filter Pack
The k-Edge Mammo filter pack is used to determine the
correction factor (recalibrate) for the Linear Mammo filter
pack.
Proceed as follows to do the recalibration:
1
Select ■ k-Edge Mammo… with
the arrow keys, then accept.
The screen will show the present
k-Edge calibration values.
2
Accept ■ Recalibrate… to start
the calibration.
You will see the calibration startup
screen as shown below.
3
Follow the screen instructions.
4
Close the instruction banner and
start the calibration.
You will be guided through the calibration procedure by
screen messages. See also ‘Cadmium K-edge Mammo
Filter Pack Calibration Overview’ on page 2-3