Kvp scope mode presets – Fluke Biomedical 199XRAY User Manual

Page 20

Advertising
background image

Fluke 199XRAY
Users Manual

2-12

kVp Scope Mode Presets

When entering the kVp mode some Scope functions are
preset. Some of these function setting are fixed, some can
be changed. See Table 2-1 and Table 2-2 for details.

Trigger settings can not be changed.

Vertical defection can be changed using the RANGE
key.

Time base can be changed using the TIME key. The
highest time base speed is 500 μs/div.

Sensitivity, Probe Factor, and Bandwidth filter are
fixed.

Note:

In kVp mode Input B can be used , but will show
no readings. A trace will only be shown if
triggering via Input A occurs.

Table 2-1. Default Filter Pack Settings

Filter Pack

Vertical Sensitivity

Zero Value

Horizontal Deflection

Trigger Level

Wide Range

20 kV/div

40 kV

10 ms/div

80 kV

K-Edge Mammo

500 mV/div

0 V

10 ms/div

1 V

Linear Mammo

5 kV/div

0 kV

10 ms/div

10 kV

CT

20 kV/div

60 kV

10 ms/div

100 kV

Mobile Pack Plus

20 kV/div

40 kV

10 ms/div

80 kV

Low Range

20 kV/div

26 kV

10 ms/div

46 kV