Fluke Biomedical 199XRAY User Manual

Page 21

Advertising
background image

Using the X-RAY kVp Function

kVp Scope Mode

2

2-13

Table 2-2. kV Range and Trigger Level vs. Filter Pack and Vertical Sensitivity

Filter Pack

Vertical Sensitivity

kV Range

Trigger Level

Wide Range

20 kV/div *

80 -150 kV

80 kV

10 kV/div

60 – 100 kV

60 kV

5 kV/div

50 – 70 kV

50 kV

K-Edge Mammo

500 mV/div*

N/A

1 V

Linear Mammo

5 kV/div *

10 – 30 kV

10 kV

10 kV/div

20 – 40 kV

20 kV

CT

20 kV/div*

100 – 140 kV

100 kV

10 kV/div

80 -120 kV

80 kV

5 kV/div

70 – 90 kV

70 kV

Mobile Pack Plus

20 kV/div*

80 – 135 kV

80 kV

10 kV/div

60 – 100 kV

60 kV

5 kV/div

50 – 70 kV

50 kV

Low Range

10 kV/div*

46 – 90 kV

46 kV

5 kV/div

36 – 60 kV

36 kV

2 kV/div

30 – 40 kV

30 kV

* Default settings.