Linear mammo filter pack – Fluke Biomedical 199XRAY User Manual

Page 14

Advertising
background image

Fluke 199XRAY
Users Manual

2-6

Linear Mammo Filter Pack

Proceed as follows to select the filter pack:

1

Select ■ Linear Mammo… with
the arrow keys, then accept.

2

Select and accept:

Recalibrate… to determine the
correction factor for the linear
Mammo filter pack.
Continue as for the k-Edge
Mammo filter pack step 3.

Use cal. values… to use the
present calibration values.

The test tool will enter the kVp
Scope Mode, see page 2-7. This
choice is disabled if the calibration
values are invalid. Select ■
Recalibrate… to get valid
calibration values again.

No correction… to perform
kVp measurements without
applying the correction factors for
the filter pack.

The test tool will enter the kVp
Scope Mode, see page 2-7.