Linear mammo filter pack – Fluke Biomedical 199XRAY User Manual
Page 14
Advertising

Fluke 199XRAY
Users Manual
2-6
Linear Mammo Filter Pack
Proceed as follows to select the filter pack:
1
Select ■ Linear Mammo… with
the arrow keys, then accept.
2
Select and accept:
■ Recalibrate… to determine the
correction factor for the linear
Mammo filter pack.
Continue as for the k-Edge
Mammo filter pack step 3.
■ Use cal. values… to use the
present calibration values.
The test tool will enter the kVp
Scope Mode, see page 2-7. This
choice is disabled if the calibration
values are invalid. Select ■
Recalibrate… to get valid
calibration values again.
■ No correction… to perform
kVp measurements without
applying the correction factors for
the filter pack.
The test tool will enter the kVp
Scope Mode, see page 2-7.