Specifications – Fluke Biomedical 199XRAY User Manual

Page 22

Advertising
background image

Fluke 199XRAY
Users Manual

2-14

Specifications

The 199XRAY kVp Measurement Accuracy Specifications
are shown in table 2-3.

Table 2-3. 199XRAY kVp Measurement Accurcy Specifications

kVp-Avg Accuracy

Filter Pack

Large Signal

Small Signal

kV Peak Accuracy*

Exposure Time Accuracy

Wide-Range 1%

1.5%

2%

0.5%

K-Edge Mammo

1%

N/A

N/A

0.5%

Linear Mammo

1%

1.5%

2%

0.5%

CT 1%

1.5%

2%

0.5%

Mobile Pack Plus**

1%

2%

2%

0.5%

Low Range

1%

1.5%

2%

0.5%

* The kV peak measurement accuracy may be affected by measurement system noise and waveform artifacts.

**Maximum error with 100% ripple