Kvp measurement procedure, Selecting a filter pack, Using the x-ray kvp function – Fluke Biomedical 199XRAY User Manual

Page 9: Chapter 2

Advertising
background image

2-1

Chapter 2

Using the X-RAY kVp Function

kVp Measurement Procedure

To perform kVp measurements, do the following:

1

Follow the operating instructions for the filter
pack as described in the filter pack Operators
Manual.

Connect the kVp divider to Input A of the
199XRAY using a shielded 1:1 cable, for
example the cable of the MA190 Accessory set.

2

Select the required filter pack, see ‘Selecting a
Filter Pack’.

3

Start the measurement. See kVp Scope Mode
on page 2-7 for further instructions.

Selecting a Filter Pack

To select a filter pack, do the following:

1

Display the
MEDICAL/RECORDER screen.

2

Select ■ kVp with the arrow keys,
then accept with the F4 (ENTER)
key. One of the Filter Pack
Selection Screens will be shown.