Using the X-RAY kVp Function

kVp Scope Mode

2

2-11

F1…F4 Softkey Functions

F1…F4 softkey functions for cursor menu OFF:

The key label shows the current filter pack
name.

Press to return to the Filter pack selection
screen if you want to select another filter pack.

Select optimal scope settings for RADIO or
FLUORO mode measurements (if applicable).

Select between TIME (in ms) or PULSE#
reading in the RADIO mode. PULSE# is not
available when HF/Three Phase signal shape
is selected.

Exit the kVp mode and return to the
MEDICAL/RECORDER screen.

F1…F4 softkey functions for cursor menu ON:

Select the cursor mode:

AUTO-kVp mode

MANUAL mode: Two vertical cursors. Use

to move the cursors. Use F2 to select

cursor to be moved.

kVp measurements apply to the area between
the cursors.

MANUAL mode: two horizontal cursors. Use

to move the cursors. Use F2 to

select cursor to be moved.

Readings are:
- voltage difference between cursors (

A

)

- voltage at the upper (

HIGH

) and lower

(

LOW

) cursor position

MANUAL MODE: One vertical cursor. Use

to move the cursor.

Readings are:
- the average, minimum, and maximum

voltage at the cursor position

Select the cursor to be moved.

Select between TIME or PULSE# reading.

Exit the MANUAL mode and return to the
AUTO kVp mode.