8. What if …

This chapter will help you to recognize malfunctions and take
appropriate measures or, if possible and acceptable, to perform
some simple repairs.

8.1 Error messages

The furnace continuously checks all functions during operation.
If an error is detected, the respective error message is
displayed.

In case of an error, the heater switches off for safety reasons.

The following error messages may be displayed:

Index Category Error

ERR Conti-

Error Message Text

No. nuation

possible


1

Entry

T < B

2

Enter a logical value for T

2

Entry

L > T

8

Enter a logical value for long-term cooling L

3

Entry

V2x <= V1x

9

Enter a logical value for the vacuum-on temperature Vx1 or the vacuum-off temperature Vx2


4

Entry

V2x > Tx + 1°C

10

Change either the vacuum values or the holding time T

5

Entry

Incorrect values for V1x, V2x 11

Enter a logical value for V1x, V2x

6

System Current temperature after 13

Excess temperature! Program aborted, furnace head opens to allow the furnace to cool down.

Start > Tx + 50 °C

*, **

8

Entry

T2 < T1

16

Enter a lower value for T1 or a higher value for T2.

9

System Power failure > 10 s during 17

A firing program in progress was interrupted for more than 10 s. The program cannot be continued!

a firing program in progress

10

Entry

T1 > V12

18

Enter a lower value for T1 or a higher value for V12


11

Entry

vV set, but V2 is

19

Pre-vacuum activated! V2 must be higher than B.

missing or invalid

12

System Error in the heating system 20

no

Check the heater fuse. If the fuse is O.K., contact your local Ivoclar Vivadent Service Center.

**

13

System Heating muffle very old

23

The heating muffle is very old. It is recommended to replace it. After the error message has been

acknowledged, a firing program may still be started.

14

System Heating muffle defective

24

The condition of the muffle is so poor that is has to be replaced immediately.


16

Entry

T is > B + 200 °C at the

26

Firing chamber too hot to start a firing program.

start of a firing program

17

System Furnace head cannot

27

The furnace head cannot be moved to the final position. It might be blocked by an external mechanical

be initialized

**, ***

source! If this is not the case, please contact your local Ivoclar Vivadent Service Center!

18

System The furnace head does not 28

The furnace head does not open/close correctly. The furnace head was manually moved or is obstructed.

reach the target position

**

The furnace head must only be moved using the keys intended for this purpose!


21

System Necessary vacuum (xxxmbar) 33

The vacuum cannot be established. Check the seal of the firing chamber, vacuum hose, vacuum pump,

is not reached within 1 min.

pump fuse.

33

Entry

HV > H (H2)

110

Enter a lower value for HV or a higher value for H (H2)


35

Entry

„Share of the holding time 120

Activate the vacuum during the holding time Tx or deactivate HV.

with vacuum“ is activated,

but Vx2 does not

correspond to Tx or Tx+1

38

System Brief power failure during a 702

A firing program in progress was interrupted by a brief power failure. The program is continued!

firing program in progress

45

System Vacuum drop

801

An unacceptable vacuum drop has occurred.

46

System The vacuum does not

802

No vacuum increase could be measured. Check the following points: Is the firing chamber tight (no

increase (self-test)

contamination on the sealing surfaces)? Is the vacuum hose connected? Is the vacuum pump connected?

Is the fuse F1 o.k.?


75

System ATK2 calibration:

1302

Error during calibration. Sample may not be correctly inserted. Try again with a new sample and make sure

Pre-heating to 963 °C

**

the sample makes ampl contact.

79

Note

Calibration reminder

1310

Some time has passed since the last calibration procedure. Calibrate the furnace soon.

80

Note

Dehumidification reminder

1312

Some time has passed since the last dehumidification procedure. Dehumidificat the furnace soon.

Behaviour of the furnace in case of error
*

Furnace head opens when this error occurs.

** A program in progress is stopped.
*** The error cannot be acknowledged; the programs cannot be started.