Product specifications – Ivoclar Vivadent P300 User Manual

Page 30

Advertising
background image

9. Product Specifications

9.1 Delivery form

– Programat P300/G2
– Power cord
– USB data cable
– Vacuum hose
– Calibration Test Pack
– Programat Firing Tray Kit
– Operating Instructions
– Service Passport Equipment

9.1.2 Recommended accessories
– Programat Accessories Set
– Temperature Checking Set 2
– Vacuum Pump VP4 / VP3 easy

9.2 Technical data

Power supply

110–120 V / 50–60 Hz
200–240 V / 50–60 Hz

Overvoltage category II
Contamination level 2

Tolerated voltage fluctuations

+/- 10%

Max. power consumption

12 A at 110–120 V

8.5 A at 200–240 V

Acceptable data for vacuum pump of other manufacturers
Max. output:

250 W / max. leakage current 0.75 mA

Final vacuum:

< 50 mbar

Use only tested pumps

Electrical fuses:

110–120 V:

250 V / T 15 A (heating circuit)

250 V / T 5 A (vacuum pump)

200–240 V:

250 V / T 8 A (heating circuit)

250 V / T 3.15 A (vacuum pump)

Dimensions of electrical fuses

110–120 V:

Diameter 6.3 x 32 mm

200–240 V:

Diameter 5 x 20 mm

Dimensions of the closed furnace

Depth: 430 mm / width: 305 mm / 410 mm (with Cooling tray)

Height: 320 mm

Usable size of the firing chamber

Diameter 80 mm

Height 48 mm

Max. firing temperature

1200 °C (2192 °F)

Weight

Furnace base: 12.0 kg

Furnace head: 4.5 kg

Safety information
The P300/G2 complies with the following guidelines:
– IEC 1010-1/EN 61010, Part 1
– UL and cUL standards

Radio protection / electromagnetic compatibility

EMC tested

9.3 Acceptable operating conditions

Acceptable ambient temperature range:
+5 °C to +40 °C (+41 °F to +104 °F)

Acceptable humidity range:
80 % maximum relative humidity for temperatures up to 31 °C
(87.8 °F) gradually decreasing to 50 % relative humidity at 40 °C
(104 °F); condensation excluded.

Acceptable ambient pressure:
The furnace is tested for use at altitudes of up to 2000 m above sea
level.

9.4 Acceptable transportation and storage conditions

Acceptable temperature range

-20 °C to +65 °C (-4 °F to +149 °F)

Acceptable humidity range

Max. 80 % relative humidity

Acceptable ambient pressure

500 mbar to 1060 mbar

Use only original packaging of the Programat P300/G2 together with
the respective foam material for shipping purposes.