Step-by-step application procedure

1. Thoroughly clean the enamel surface
to be sealed.

2. Isolate the working field, preferably
with a rubber dam.

3. Apply an etching gel, e.g. Email
Preparator, and let it react for 30 to
60 seconds.

4. Rinse thoroughly.

5. Dry with water- and oil-free air. The
etched enamel should have a mat white
appearance. Avoid contamination of the
etched surface with saliva.

6. Apply Helioseal F directly with the
disposable cannula or a disposable
brush, and disperse.

7. Wait for approx. 15 seconds. Then cure
the sealant with a suitable poly merization
light (e.g. bluephase

®

) for 20 seconds.

8. Check seal and occlusion.

Warning
Avoid contact of unpolymerized material with
skin/mucous membrane or eyes. Unpolymerized
Helioseal F may cause slight irritation and, in rare
cases, may lead to a sensitization against
methacrylates. Commercial medical gloves do not

provide protection against the sensitizing effect of
methacrylates.

Special note

If Helioseal F is applied from the Cavifil

directly in the mouth of the patient, we rec-

ommend using this Cavifil only for one patient due
to hygienic reasons (prevention of cross-contamina-
tion between patients). The same applies to the
application tips of the syringe.

Storage
– Close Helioseal F syringe immediately after use
– Store material at 2–28 °C (36–82 °F).
– Shelf life: see date of expiration.
– Do not use the material after the indicated date

of expiration.

– Syringes or Cavifils should not be disinfected

with oxidizing disinfection agents.

Keep out of the reach of children!
For use in dentistry only!

Date information prepared
02/2010, Rev. 2

Manufacturer
Ivoclar Vivadent AG
FL-9494 Schaan / Liechtenstein

The material has been developed solely for use in dentistry. Processing should be carried out
strictly according to the Instructions for Use. Liability cannot be accepted for damages
resulting from failure to observe the Instructions or the stipulated area of application. The
user is responsible for testing the material for its suitability and use for any purpose not
explicitly stated in the Instructions. Descriptions and data constitute no warranty of attribut-
es and are not binding.