PR-03 instruction manual

8.4.6 PR-03-P programming specifics

If the optical image appears inverted when using the probe refractometer PR-03-P (see Figure 8.12), con-
figure parameter lminv to value 1 by key-in sequence Calibrate / Parameters / Standard RI
(25

C) / Image inversion

to turn the picture right way.

A

B

C

D

Slope

TEST: 248.0

30.2 ºC

Endp: 11

L): -7.5

R(:

3.3

HT: 20 ºC

HH: 4

%

SCALED

OPTICAL

IMAGE

Low concentration / No sample

Figure 8.12

Inverted optical image

Note:

This image inversion correction is only available in program version 8.5 or higher. When an IT-R

is delivered together with a Probe refractometer sensor, the program version is always compatible with the
inverted image.