Aa1l(data), Aa1l(data) -16 – Measurement Computing CB-7080 User Manual

Page 34

Advertising
background image

CB-7080 & CB-7080D Counter/Timer User's Guide

$AA1L(data)

2-16

$AA1L(data)

Description

: Set the low trigger level of non-isolated input. Refer to "

Programmable threshold voltage

setting

" for more information.

Syntax

: $AA1L(data)[chk](cr)

$

A delimiter character

AA

2-character HEX module address, from 00 to FF

(data)

2-character decimal value for low trigger level. The unit is 0.1V and the

range can be from 0.0 to 5.0V

[chk]

2-character checksum. If checksum is disabled, no [chk].

(cr)

0x0D

Response

:

Valid response

!AA[chk](cr)

Invalid response

?AA[chk](cr)

no response

Syntax error, communication error or address error

!

A delimiter character indicating a valid response

?

A delimiter character indicating an invalid response

AA

2-character HEX module address

[chk]

2-character checksum. If checksum is disabled, no [chk].

(cr)

0x0D

Examples

:

Low trigger level=0.8 volt:

command:

$011L08(cr)

response:

!01(cr)

Low trigger level=1.0 volt:

command:

$021L10(cr)

response:

!02(cr)

Note:

The default value for the low trigger level is 0.8V.