8-Advanced Function of Oscilloscope

Now, you can see a screen that looks like the following figure 51.

figure 51: Use the Cursor for a Time Measurement

8.12 Autoscale

The function is applied to follow-up signals automatically even if the signals change at any time.
Autoscale enables the instrument to set up trigger mode, voltage division and time scale
automatically according to the type, amplitude and frequency of the signals.
The menu is as follows:

Function

menus

Settings Description

Autoscale

OFF

ON

Turn off Autoscale

Turn on Autoscale

Mode

Vertical

Horizontal

HORI—VERT

Only adjust vertical scale according to input signal.

Only adjust horizontal scale according to input signal.

Adjust both vertical and horizontal scales according to

input signal.

Display only one or two cycles in waveform.

Display Multi-cycles in waveform.

If you want to measure voltage of Channel 1, you can do as the follows:

1. Press MENU, the function menu will appear on the right of the screen.

2. Press MENU or MENU and choose Autoscale, three options will show at the