7.SCPI

40

7. SCPI

The waveform generator supports SCPI, and the users can operate and control the device by USB port.
For the detailed information about SCPI, please refer to

AG Series Waveform Generator SCPI

Protocol.

8. Troubleshooting

1. The instrument is powered on but no Display.

Check if the power is connected properly.

Check if the Power Switch is in the proper voltage scale.

Check if the fuse which is below the AC Power socket is used appropriately and
in good condition (the cover can be pried open with a straight screwdriver).

Restart the instrument after the steps above.

If the problem still exists, please contact Lilliput for our service.

2. The measured value of output signal amplitude disaccords to the displayed value:

Check if the actual load matches the set one. Please refer to To Set the Output Load

on page 35.

If you encounter other problems, try to reset the settings (refer to To Return to Default

Setting on page 36) or restart the instrument. If it still can not work properly, please

contact Lilliput for our service.