Seaward DO7010 User Manual
Page 29

4-1-1 Measurement range selection:
Φ Select the RANGE line
Φ Press on the RIGHT arrow or on the enter key .
6000m
Ω : from 0001 to 6000 mΩ
600m
Ω : from 000.1 to 600.0 mΩ
60m
Ω : from 00.01 to 60.00 mΩ
6m
Ω : from 0.001 to 6.000 mΩ
Φ Enter the selection with the Enter key.
4-1-2 Measurement current selection:
Φ Select the CURRENT line
Φ Press on the RIGHT arrow or on the enter key.
Φ Select the measurement current with the UP and DOWN arrows:
HI = 10A for the 6 to 600m
Ω ranges and 1A for the 6000 mΩ range.
LO = 1A for the 6 to 600m
Ω and 0.1A for the 6000 mΩ range.
Φ Enter the selection with the Enter key.
RANGE LO
CURRENT
>
HI
ZERO
AVERAGE
LOG
LIMITS
ESC
RANGE
>
6000mΩ
CURRENT 600mΩ
ZERO
60mΩ
AVERAGE 6mΩ
LOG AUTO1
LIMITS
AUTO2
ESC
Φ Select the measurement range with the UP and DOWN arrows:
AUTO2 : automatic searching range from the last measurement range.
AUTO1 : automatic searching range from the highest range (6000m ).
Ω
Operating Instructions
Digital Microhmmeter Type DO7010
27
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•