Operating Instructions

- 25 -

8 Electrical Specifications

8.1 Earth Resistance Two/Three Pole Method

Display Range 0.00W to 50.0kW
Measurement
Ranges

0Ω - 19.99Ω

Resolution 0.01Ω

20Ω - 199.9Ω

Resolution 0.1Ω

200Ω - 199.9Ω

Resolution 1Ω

1.00KΩ - 9.99KΩ Resolution 10Ω

10.0KΩ - 50.0KΩ Resolution 100Ω

Accuracy

2% ± 2 counts from 0Ω - 999Ω

5% ± 2 counts from 1.00kΩ - 50.0kΩ

Measuring
frequency

125 Hz ± 1HZ

Measuring
current

<20mA effective

Measuring

voltage

<30V effective sine wave

Distortion

Max 5V p-p / 50Hz


8.2 Earth Resistance Current Clamp Method

Display Range 0.00W to 2.0kW
Measurement
Ranges

0.5Ω - 19.9Ω Resolution 0.1Ω

20Ω – 999Ω

Resolution 1Ω

1.00Ω - 2.00kΩ Resolution 10Ω

Accuracy

5% + 2 counts

Measuring
frequency

125 Hz ± 1Hz

Measuring

current

<20mA effective

Measuring

voltage

<30V effective sine wave

Distortion

Max 5V p-p / 50Hz

Note: the stated accuracy of the earth resistance

current clamp method does not include errors
introduced by the current clamp