Seaward ARE 1557 User Manual

Page 27

Advertising
background image

ERT1557

Operating Instructions

- 27 -

9 Environmental Conditions

9.1 The ERT 1557 has been designed to perform tests

and measurements in a dry environment.


9.2 Maximum barometric elevation for making

measurements is 2000M.

9.3 Overvoltage category IEC 60664/IEC 61010, 300V

Category II.


9.4 Pollution degree 2 according to IEC 61010-1.

9.5 Protective system IP40 according to IEC 60529.

9.6 Electromagnetic compatibility (EMC). Interference

immunity and emitted interference conforming to
IEC 61326-1.


9.7 Operating temperature range of 0°C to 40°C,

without moisture condensation.

9.8 The ERT 1557 can be stored at any temperature in

the range -25°C to +65°C (relative humidity up to
90%). The batteries should be taken out of the

instrument for storage.

9.9 Operating Altitude 0 to 2000 metres

Advertising